    In Vitro Evaluation of ECG-Synchronized Pulsatile Flow Using the i-cor Diagonal Pump in Neonatal and Pediatric ECLS Systems.

    • Morgan Moroi, Madison Force, Shigang Wang, Allen R Kunselman, and Akif Ündar.
    • Department of Pediatrics, Penn State Health Pediatric Cardiovascular Research Center, Penn State College of Medicine, Penn State Health Children's Hospital, Hershey, PA, USA.
    • Artif Organs. 2018 Jul 1; 42 (7): E127-E140.

    AbstractThe objective was to assess the i-cor electrocardiogram-synchronized diagonal pump in terms of hemodynamic energy properties for off-label use in neonatal and pediatric extracorporeal life support (ECLS) circuits. The neonatal circuit consisted of an i-cor pump and console, a Medos Hilite 800 LT oxygenator, an 8Fr arterial cannula, a 10Fr venous cannula, 91 cm of 0.6-cm ID arterial tubing, and 91 cm of 0.6-cm ID venous tubing. The pediatric circuit was identical except it included a 12Fr arterial cannula, a 14Fr venous cannula, and a Medos Hilite 2400 LT oxygenator. Neonatal trials were conducted at 36°C with hematocrit 40% using varying flow rates (200-600 mL/min, 200 mL increments) and postarterial cannula pressures (40-100 mm Hg, 20 mm Hg increments) under nonpulsatile mode and pulsatile mode with various pulsatile amplitudes (1000-4000 rpm, 1000 rpm increments). Pediatric trials were conducted at different flow rates (800-1600 mL/min, 400 mL/min increments). Mean pressure and energy equivalent pressure increased with increasing postarterial cannula pressure, flow rate, and pulsatile amplitude. Physiologic-like pulsatility was achieved between pulsatile amplitudes of 2000-3000 rpm. Pressure drops were greatest across the arterial cannula. Pulsatile flow generated significantly higher total hemodynamic energy (THE) levels than nonpulsatile flow. THE levels at postarterial cannula site increased with increasing postarterial cannula pressure, pulsatile amplitude, and flow rate. No surplus hemodynamic energy (SHE) was generated under nonpulsatile mode. Under pulsatile mode, preoxygenator SHE increased with increasing postarterial cannula pressure and pulsatile amplitude, but decreased with increasing flow rate. The i-cor system can provide nonpulsatile and pulsatile flow for neonatal and pediatric ECLS. Pulsatile amplitudes of 2000-3000 rpm are recommended for use in neonatal and pediatric patients.© 2018 International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…