    Hepatic bipolar radiofrequency ablation using perfused-cooled electrodes: a comparative study in the ex vivo bovine liver.

    • J M Lee, J K Han, S H Kim, J Y Lee, S H Choi, and B I Choi.
    • Department of Radiology and Institute of Radiation Medicine, Seoul National University College of Medicine and Clinical Research Institute, Seoul National University Hospital, Seoul 110-744, Korea.
    • Br J Radiol. 2004 Nov 1; 77 (923): 944-9.

    AbstractThe purpose of this paper was to demonstrate the efficacy of the dual probe bipolar radiofrequency (RF) system with the perfused-cooled electrodes inducing coagulation necrosis in the ex vivo bovine liver. The perfused-cooled electrode that allows simultaneous internal cooling and interstitial hypertonic saline perfusion has been developed for RF ablation (RFA). RF was applied to excised bovine liver in a bipolar mode at 150 W using a 200 W generator with two perfused-cooled electrodes for 10 min. After placing the electrodes at 3 cm spacing in the explanted liver, 45 ablation zones were created with three different regimens: Group A, using both intraelectrode cooling and interstitial perfusion; group B, using only the intraelectrode cooling; and group C, using only interstitial perfusion. In groups A and C, RFA was performed with the infusion of 6% hypertonic saline at the rate of 2 ml min(-1). During RFA, we measured the tissue temperature at the midpoint between the two electrodes. The dimensions of the ablation zones and the changes in impedance, currents and liver temperature during RFA were compared in these three groups. The mean tissue impedance during RFA in group A (56.7+/-21.7 Omega) and group C (56.9+/-20.6 Omega) was significantly lower than group B (112+/-19.7 Omega) (p<0.001). The mean current was higher in group A (1765+/-128 mA) than groups B (760+/-321 mA) and C (1298+/-349 mA) (p<0.05). In addition, the shortest vertical diameter of coagulation necrosis was greater in groups A (4.9+/-0.5 cm) and C (4.6+/-0.7 cm) than in group B (3.5+/-0.4 cm) (p<0.05). The temperature at the mid-point between the two probes was higher in group A than other groups: 99 degrees C in group A, 88.9 degrees C in group B, and 94.3 degrees C in group C (p>0.05). The ratios of the diameter of the long-axis to the diameter of the vertical-axis of groups A, B and C were 1.1+/-0.1, 1.2+/-0.1, and 1.1+/-0.2, respectively (p<0.05). Bipolar RFA using intraelectrode cooling and the interstitial saline perfusion simultaneously produced ablation zones significantly larger than the area produced by only one measure.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…