- Max Wacker, Ingo Slottosch, Maximilian Scherner, Sam Varghese, and Jens Wippermann.
- Department of Cardiothoracic Surgery, University Hospital of Magdeburg, Magdeburg, Germany. Electronic address: max.wacker@med.ovgu.de.
- Ann. Thorac. Surg. 2019 Apr 1; 107 (4): e243-e244.
AbstractOver the last decade, aortic valve replacement with rapid-deployment biologic bioprostheses has become a common alternative to the use of mechanical or biologic stented valves for high-risk patients. A 63-year-old patient underwent uncomplicated valve replacement with the Edwards Intuity valve (Edwards Lifesciences, Irvine, CA). Two months postoperatively, the patient had progressive dyspnea. Echocardiography showed a dislocated aortic valve reaching into the left ventricular outflow tract and impairing the anterior mitral valve leaflet. Both valves were replaced by a conventional stented bioprosthesis. This case report describes the first case of valve migration for the Edwards Intuity Elite rapid-deployment aortic valve and discusses possible explanations.Copyright © 2019 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.