• Spine J · Oct 2018

    The effect of interbody fusion cage design on the stability of the instrumented spine in response to cyclic loading: an experimental study.

    • Ron N Alkalay, Robert Adamson, and Michael W Groff.
    • Center for Advanced Orthopaedic Studies, Department of Orthopaedics, Beth Israel Deaconess Medical Center and Harvard Medical School, 330 Brookline Ave, Boston, MA 02215, USA. Electronic address: ralklay@bidmc.harvard.edu.
    • Spine J. 2018 Oct 1; 18 (10): 1867-1876.

    Background ContextIn the lumbar spine, end plate preparation for the interbody fusion cages may critically affect the cage's long-term performance. This study investigated the effect of the interbody cage design on the compliance and cage subsidence of instrumented spines under cyclic compression.PurposeWe aimed to quantify the role of cage geometry and bone density on the stability of the spinal construct in response to cyclic compressive loads.Study DesignChanges in the cage-bone interface and the effect of bone density on these changes were evaluated in a human cadaveric model for three intervertebral cage designs.MethodsThe intervertebral space of 27 functional cadaveric spinal units was instrumented with bilateral linear cages, single anterior conformal cages, or single unilateral oblique cages. Once augmented with a pedicle screw fixation system, the instrumented spine unit was tested under cyclic compression loads (400-1,200 N) to 20,000 cycles at a rate of 2 Hz. Compliance of the cage-bone interface and cage subsidence was computed. Two-way repeated multivariate analysis of variance was used to test the effects of cage design and bone density on the compliance and subsidence of the cages.ResultsThe anterior conformal shaped cage showed reduced interface stiffness (p<.01) and higher hysteresis (p<.01) and subsidence rate (10%-30%) than the bilateral linear and unilateral oblique-shaped cages. Bone density was not associated with the initial compliance of the cage-bone interface or the rate of cage subsidence. Higher bone density did decrease the rate of reduction in cage-bone interface stiffness under higher cyclic loads for the anterior conformal shaped and unilateral oblique cages.ConclusionsCage design and position significantly affected the degradation of the cage-bone interface under cyclic loading. Comparisons of subsidence rate between the different cage designs suggest the peripheral location of the cages, using the stronger peripheral subchondral bone of the apophyseal ring, to be advantageous in preventing the subsidence and failure of the cage-bone interface.Copyright © 2018. Published by Elsevier Inc.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…