• Medicine · Sep 2021

    Observational Study

    Removal of a catheter mount and heat-and-moisture exchanger improves hypercapnia in patients with acute respiratory distress syndrome: A retrospective observational study.

    • Takaya Shimoda, Motohiro Sekino, Ushio Higashijima, Sojiro Matsumoto, Shuntaro Sato, Rintaro Yano, Takashi Egashira, Hiroshi Araki, Iwasaki Naoya, Suzumura Miki, Ryo Koyanagi, Makoto Hayashi, Shintaro Kurihara, and Tetsuya Hara.
    • Medical Engineering Equipment Center, Nagasaki University Hospital, Nagasaki, Japan.
    • Medicine (Baltimore). 2021 Sep 10; 100 (36): e27199e27199.

    AbstractTo avoid ventilator-associated lung injury in acute respiratory distress syndrome (ARDS) treatment, respiratory management should be performed at a low tidal volume of 6 to 8 mL/kg and plateau pressure of ≤30 cmH2O. However, such lung-protective ventilation often results in hypercapnia, which is a risk factor for poor outcomes. The purpose of this study was to retrospectively evaluate the effectiveness and safety of the removal of a catheter mount (CM) and using heated humidifiers (HH) instead of a heat-and-moisture exchanger (HME) for reducing the mechanical dead space created by the CM and HME, which may improve hypercapnia in patients with ARDS.This retrospective observational study included adult patients with ARDS, who developed hypercapnia (PaCO2 > 45 mm Hg) during mechanical ventilation, with target tidal volumes between 6 and 8 mL/kg and a plateau pressure of ≤30 cmH2O, and underwent stepwise removal of CM and HME (replaced with HH). The PaCO2 values were measured at 3 points: ventilator circuit with CM and HME (CM + HME) use, with HME (HME), and with HH (HH), and the overall number of accidental extubations was evaluated. Ventilator values (tidal volume, respiratory rate, minutes volume) were evaluated at the same points.A total of 21 patients with mild-to-moderate ARDS who were treated under deep sedation were included. The values of PaCO2 at HME (52.7 ± 7.4 mm Hg, P < .0001) and HH (46.3 ± 6.8 mm Hg, P < .0001) were significantly lower than those at CM + HME (55.9 ± 7.9 mm Hg). Measured ventilator values were similar at CM + HME, HME, and HH. There were no cases of reintubation due to accidental extubation after the removal of CM.The removal of CM and HME reduced PaCO2 values without changing the ventilator settings in deeply sedated patients with mild-to-moderate ARDS on lung-protective ventilation. Caution should be exercised, as the removal of a CM may result in circuit disconnection or accidental extubation. Nevertheless, this intervention may improve hypercapnia and promote lung-protective ventilation.Copyright © 2021 the Author(s). Published by Wolters Kluwer Health, Inc.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.