Investigation and Analysis of Hemoglobin A1c Measurement Systems' Performance for 135 Laboratories in China.
- Hai-Jian Zhao, Tian-Jiao Zhang, Jie Zeng, Cui-Hua Hu, Rong Ma, and Chuan-Bao Zhang.
- National Center for Clinical Laboratories, Beijing Hospital, National Center of Gerontology, Beijing Engineering Research Center of Laboratory Medicine, Beijing 100730, China.
- Chin. Med. J. 2017 May 5; 130 (9): 107910841079-1084.
BackgroundHemoglobin A1c (HbA1c) measurement is of great value for the diagnosis and monitoring of diabetes. Many manufacturers have developed various experiments to determine the HbA1c concentration. However, the longitudinal use of these tests requires strict quality management. This study aimed to analyze the quality of HbA1c measurement systems in China using six sigma techniques to help improve their performances.MethodsA total of 135 laboratories were involved in this investigation in 2015. Bias values and coefficients of variation were collected from an HbA1c trueness verification external quality assessment program and an internal quality control program organized by the National Center of Clinical Laboratories in China. The sigma (σ) values and the quality goal index (QGI) were used to evaluate the performances of different groups, which were divided according to principles and instruments.ResultsThe majority of participants (88, 65.2%) were scored as "improvement needed (σ < 3)", suggesting that the laboratories needed to improve their measurement performance. Only 8.2% (11/135) of the laboratories were scored as "world class (σ ≥ 6)". Among all the 88 laboratories whose σ values were below 3, 52 (59.1%) and 23 (26.1%) laboratories needed to improve measurement precision (QGI <8.0) and trueness (QGI >1.2), respectively; the remaining laboratories (13, 14.8%) needed to improve both measurement precision and trueness. In addition, 16.1% (5/31) and 15.0% (3/20) of the laboratories in "TOSOH" and "ARKRAY" groups, respectively, were scored as "world class", whereas none of the laboratories in "BIO-RAD" group were scored as "world class".ConclusionsThis study indicated that, although participating laboratories were laboratories with better performance in China, the performances were still unsatisfactory. Actions should be taken to improve HbA1c measurement performance before we can include HbA1c assays in diabetes diagnosis in China.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.