• Acta Anaesthesiol Scand · Jan 2023

    Glycemic control and blood gas sampling frequency during continuous glucose monitoring in the ICU: A before-and-after study.

    • Johan Mårtensson, Salvatore Cutuli, Fumitaka Yanase, Paolo Ancona, Lisa Toh, Eduardo Osawa, and Rinaldo Bellomo.
    • Department of Physiology and Pharmacology, Section of Anaesthesia and Intensive Care, Karolinska Institutet, Stockholm, Sweden.
    • Acta Anaesthesiol Scand. 2023 Jan 1; 67 (1): 869386-93.

    BackgroundWhether subcutaneous continuous glucose monitoring (CGM) can safely replace intermittent arterial blood gas glucose analyses in intensive care unit (ICU) patients remains uncertain. We aimed to compare CGM to blood gas glucose values and assess whether CGM use reduces blood gas sampling frequency and glucose variability in ICU patients with type 2 diabetes managed with liberal glucose control.MethodsWe used the FreeStyle Libre CGM in 15 ICU patients and compared their blood glucose metrics with a pre-CGM control population of 105 ICU patients with type 2 diabetes. Both groups received insulin to target glucose range of 10-14 mmol/L. We used linear regression analysis adjusted for illness severity to assess the association of CGM use with blood gas sampling frequency and glucose variability. We used mean absolute relative difference (MARD) and Clarke error grid analysis to assess accuracy of matched CGM-blood glucose values overall, across glucose stata (<10, 10-14, >14 mmol/L), and over time (≤48, 48-96, >96 h).ResultsWe analyzed 483 matched glucose values. Overall MARD was 11.5 (95% CI, 10.7-12.3)% with 99% of readings in Clarke zones A and B. MARD was 15.5% for glucose values <10 mmol/L, 11.1% at 10-14 mmol/L, and 11.4% >14 mmol/L. MARD was 13.8% in the first 48 h, 10.9% at 48-96 h, and 8.9% beyond 96 h. CGM use was associated with 30% reduction in blood gas sampling frequency. CGM use was not associated with glucose variability as determined by glycemic lability index or standard deviation of blood glucose.ConclusionsIn our cohort of ICU patients with type 2 diabetes receiving liberal glycemic control, CGM showed acceptable accuracy and was associated with a reduction in blood gas sampling frequency without compromising glucose control. Lowest accuracy was observed at glucose values below 10 mmol/L and during the first 48 h of CGM use.© 2022 The Authors. Acta Anaesthesiologica Scandinavica published by John Wiley & Sons Ltd on behalf of Acta Anaesthesiologica Scandinavica Foundation.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…