• J Clin Monit Comput · Aug 2024

    Feasibility study of the use of a wearable vital sign patch in an intensive care unit setting.

    • Guylian Stevens, Michiel Larmuseau, Annelies Van Damme, Henk Vanoverschelde, Jan Heerman, and Pascal Verdonck.
    • Departement of Electronics and Information Systems - IBiTech, Ghent University, Korneel Heymanslaan, Gent, 9000, East-Flanders, Belgium. Guylian.Stevens@ugent.be.
    • J Clin Monit Comput. 2024 Aug 19.

    AbstractMultiple studies and review papers have concluded that early warning systems have a positive effect on clinical outcomes, patient safety and clinical performances. Despite the substantial evidence affirming the efficacy of EWS applications, persistent barriers hinder their seamless integration into clinical practice. Notably, EWS, such as the National Early Warning Score, simplify multifaceted clinical conditions into singular numerical indices, thereby risking the oversight of critical clinical indicators and nuanced fluctuations in patients' health status. Furthermore, the optimal deployment of EWS within clinical contexts remains elusive. Manual assessment of EWS parameters exacts a significant temporal toll on healthcare personnel. Addressing these impediments necessitates innovative approaches. In this regard, wearable medical technologies emerge as promising solutions capable of continual monitoring of hospitalized patients' vital signs. To overcome the barriers of the use of early warning scores, wearable medical technology has the potential to continuously monitor vital signs of hospitalised patients. However, a fundamental inquiry arises regarding the comparability of their reliability to the current used golden standards. This inquiry underscores the imperative for rigorous evaluation and validation of wearable medical technologies to ascertain their efficacy in augmenting extant clinical practices. This prospective, single-center study aimed to evaluate the accuracy of heart rate and respiratory rate measurements obtained from the Vivalink Cardiac patch in comparison to the ECG-based monitoring system utilized at AZ Maria Middelares Hospital in Ghent. Specifically, the study focused on assessing the concordance between the data obtained from the Vivalink Cardiac patch and the established ECG-based monitoring system among a cohort of ten post-surgical intensive care unit (ICU) patients. Of these patients, five were undergoing mechanical ventilation post-surgery, while the remaining five were not. The study proceeded by initially comparing the data recorded by the Vivalink Cardiac patch with that of the ECG-based monitoring system. Subsequently, the data obtained from both the Vivalink Cardiac patch and the ECG-based monitoring system were juxtaposed with the information derived from the ventilation machine, thereby providing a comprehensive analysis of the patch's performance in monitoring vital signs within the ICU setting. For heart rate, the Vivalink Cardiac patch was on average within a 5% error range of the ECG-based monitoring system during 85.11±10.81% of the measured time. For respiratory rate this was during 40.55±17.28% of the measured time. Spearman's correlation coefficient showed a very high correlation of ρ = 0.9 8 for heart rate and a moderate correlation of ρ = 0.66 for respiratory rate. In comparison with the ventilated respiratory rate (ventilation machine) the Vivalink and ECG-based monitoring system both had a moderate correlation of ρ = 0.68 . A very high correlation was found between the heart rate measured by the Vivalink Cardiac patch and that of the ECG-based monitoring system of the hospital. Concerning respiratory rate the correlation between the data from the Vivalink Cardiac patch, the ECG-based monitoring system and the ventilation machine was found to be moderate.© 2024. The Author(s), under exclusive licence to Springer Nature B.V.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…