J Vasc Interv Radiol · Nov 2004
Clinical TrialExtended interval for retrieval of Günther Tulip filters.
- Olaf Alfons Terhaar, Stuart Macalister Lyon, Mark Frank Given, Anne Elizabeth Foster, Frank Mc Grath, and Michael J Lee.
- Department of Academic Radiology, Beaumont Hospital and Royal College of Surgeons Medical School, Beaumont Road, Dublin 9, Ireland.
- J Vasc Interv Radiol. 2004 Nov 1;15(11):1257-62.
PurposeTo evaluate the Gunther Tulip vena cava filter with regard to ease of placement, complications, and retrieval over long time periods.Materials And MethodsIn 53 patients (ratio of men to women, 24:29; mean age, 52.8 years) retrievable Gunther Tulip filters (Vena Cava M Reye Filter Set; William Cook Europe, Denmark) were inserted. Indications included planned major surgery with recent pulmonary embolus or high pulmonary embolus risk (n = 16), extensive ilio-femoral thrombus (n = 11), deep vein thrombosis with anticoagulant complications (n = 9), breakthrough pulmonary embolus despite anticoagulant therapy (n = 4), and contraindication to anticoagulant therapy (n = 13). All patients were followed-up for immediate and long-term complications.ResultsFifty-three filters were successfully placed in 52 of 53 patients, yielding a success rate of 98.1%. Nineteen patients underwent attempted retrieval of their filter. Sixteen of 19 retrieval procedures were successful (84%). In three patients, the filter could not be removed on attempted retrieval (extensive filter thrombus in two patients and attachment to the wall in one patient). One patient received two filters, which were both successfully retrieved at a later date. Median implantation time for retrievable filters was 34 days (range, 7-126 days). Mean follow-up for patients with permanent filters was 13 months. Two major complications (pneumothorax and break through pulmonary embolus) and three minor complications (right internal jugular vein thrombosis in two patients and transient Horner's Syndrome in one patient) were recorded.ConclusionInsertion and retrieval of filters is safe and feasible. Preliminary data suggest that Gunther Tulip filter retrieval is feasible over and above the manufacturer's recommended retrieval interval of 14 days.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.