• J Vet Emerg Crit Care (San Antonio) · Jun 2010

    Evaluation of an indirect oscillometric blood pressure monitor in normotensive and hypotensive anesthetized dogs.

    • Andre Shih, Sheilah Robertson, Alessio Vigani, Anderson da Cunha, Luisito Pablo, and Carsten Bandt.
    • Departments of Large Animal Clinical Science, College of Veterinary Medicine, University of Florida, Gainesville, FL 32610, USA. shiha@vetmed.ufl.edu
    • J Vet Emerg Crit Care (San Antonio). 2010 Jun 1;20(3):313-8.

    ObjectiveTo determine the accuracy and precision of an oscillometric noninvasive blood pressure device as a predictor of invasive direct blood pressure in healthy anesthetized hypotensive and normotensive dogs.DesignProspective observational study.SettingUniversity teaching hospital.AnimalsEight crossbred adult dogs.InterventionsAnesthesia was induced with propofol and maintained with isoflurane. A catheter was placed in the dorsal pedal artery to record systolic, mean, and diastolic arterial blood pressures (aSAP, aMAP, and aDAP, respectively). The noninvasive blood pressure device cuff was placed around the contralateral front limb to record noninvasive systolic, mean, and diastolic blood pressure (nSAP, nMAP, and nDAP). Two states of blood pressure (BP) were studied: baseline state was established by keeping end-tidal isoflurane concentration at 1.2+/-0.1%. The hypotensive state was achieved by maintaining the same isoflurane concentration while withdrawing approximately 40% of the animal's blood volume until aMAP was stable at approximately 40 mm Hg. At the end of the study, blood was returned to the animal and it was allowed to recover from anesthesia.Measurements And Main ResultsAgreement between the direct and indirect BP measurements was determined by the Bland-Altman method. The SAP and MAP but not DAP bias varied significantly between each BP state. Normotensive absolute biases (mean [SD]) for SAP, MAP, and DAP were -14.7 mm Hg (15.5 mm Hg), -16.4 mm Hg (12.1 mm Hg), and -14.1 mm Hg (15.8 mm Hg), respectively. Absolute biases during the hypotensive state for SAP, MAP, and DAP were -32 mm Hg (22.6 mm Hg), -24.2 mm Hg (19.5 mm Hg), and -16.8 mm Hg (17.2 mm Hg), respectively.ConclusionThe oscillometric device was not reliably predictive of intra-arterial BP during hypotension associated with acute hemorrhage.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…