-
Comparative Study
Effects of Condensate in the Exhalation Limb of Neonatal Circuits on Airway Pressure During Bubble Continuous Positive Airway Pressure.
- Tiffany M Youngquist, C Peter Richardson, and Robert M Diblasi.
- Center for Developmental Therapeutics, Seattle Children's Research Institute, Seattle, Washington.
- Respir Care. 2013 Nov 1;58(11):1840-6.
BackgroundBubble CPAP is frequently used in spontaneously breathing infants with lung disease. Often bubble CPAP systems lack pressure alarms and pressure-release valves. We observed a large volume of condensate in the exhalation limb of a patient circuit and conducted a series of experiments to test the hypothesis that accumulated condensate could affect delivered pressures.MethodsAn anatomically accurate nasal airway model of a preterm infant was attached to a spontaneously breathing lung model. A bubble CPAP system was attached to the nasal airway with bi-nasal short prongs, and the rate of fluid condensation was measured. Next, tracheal pressures were monitored digitally to detect changes in airway pressure related to condensate accumulation. Measurements were obtained with volumes of 0, 5, 10, 15, and 20 mL of water in the exhalation limb, at flows of 4, 6, 8, and 10 L/min. Measurements with 20 mL in the exhalation limb were recorded with and without a pressure-relief valve in the circuit.ResultsThe rate of condensate accumulation was 3.8 mL/h. At volumes of ≥ 10 mL, noticeable alterations in the airway pressure waveforms and significant increases in mean tracheal pressure were observed. The pressure-relief valve effectively attenuated peak tracheal pressure, but only decreased mean pressure by 0.5-1.5 cm H2O.ConclusionsCondensate in the exhalation limb of the patient circuit during bubble CPAP can significantly increase pressure delivered to the patient. The back and forth movement of this fluid causes oscillations in airway pressure that are much greater than the oscillations created by gas bubbling out the exhalation tube into the water bath. We recommend continuously monitoring pressure at the nasal airway interface, placing an adjustable pressure-relief valve in the circuit, set to 5 cm H2O above the desired mean pressure, and emptying fluid from the exhalation limb every 2-3 hours.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.