• Resp Care · Aug 2012

    Case Reports

    Inadvertent triggering of the ventilator caused by surgically placed pacer wires.

    • Timothy E France, A Randolph Garnett, and David Grooms.
    • Department of Respiratory Care, Sentara Norfolk General Hospital, Norfolk, Virginia, USA. tefrance@sentara.com
    • Resp Care. 2012 Aug 1;57(8):1334-6.

    AbstractInadvertent ventilator triggering can occur for various reasons. Leaks in the ventilator circuit, endotracheal tube leaks, tracheal cuff leaks, cardiac oscillations, water condensate causing oscillations in the circuit tubing, ventilator expiratory valve integrity, and overly sensitive triggering mechanism settings may precipitate this phenomena. We present a case of inadvertent ventilator triggering caused by electrical stimulation of the diaphragm from surgically placed pacing wires post cardiothoracic surgery. A 47-year-old woman underwent surgical placement of a left (LVAD) and right (RVAD) ventricular assist device for severe end stage cardiomyopathy, as a bridge to cardiac transplantation. The patient was observed to have inadvertent ventilator triggering while deeply sedated postoperatively. The ventilator set respiratory rate was 16 breaths/min, with patient respiratory rate of 30 breaths/min while deeply sedated. Upon assessment of ventilator waveforms and arterial blood gas revealing a profound respiratory alkalosis, the pressure/time waveform demonstrated a -2 cm H(2)O decrease in pressure prior to each cycled breath. The ventilator was subsequently changed from flow trigger sensitivity of 3 L/min to pressure trigger sensitivity of -3 cm H(2)O to eliminate the autotriggering. Later in the patient's ICU stay, inadvertent ventilator triggering was again observed. Further adjustment of the pressure trigger sensitivity to -3 cm H(2)O eliminated the autotriggering. Clinical assessment found the pacing wires were responsible for stimulating the patient's diaphragm, therefore causing airway pressure decreases and premature breath delivery. Once the electrical amplitude of the pacemaker was decreased, the inadvertent ventilator triggering resolved and normal trigger sensitivity and pH was restored.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…