    Positioning artifact causing retained foreign object appearance in a radiograph of a venous catheter.

    • Arnoley S Abcejo, Hugh M Smith, and James R Hebl.
    • From the Department of Anesthesiology, Mayo Clinic College of Medicine, Rochester, Minnesota.
    • A A Case Rep. 2015 Apr 15;4(8):97-9.

    AbstractIn preparation for an abdominal tumor debulking procedure, a patient had a right internal jugular central venous catheter (CVC) inserted. A radiologist interpreted a postoperative chest radiograph as containing a 7-cm foreign body in the distal lumen of the CVC. After removal, dissection of the CVC revealed a polyurethane plug in the distal lumen. The manufacturer (Arrow®) states that the synthetic plug is part of standard CVC construction that may produce a radiopaque signal with specific patient positioning during radiography. This report is intended to raise awareness of the potential for this radiographic finding and prevent the unnecessary removal of clinically indicated CVCs.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…