Comparative Study
Causes of ventricular oversensing in implantable cardioverter-defibrillators: implications for diagnosis of lead fracture.
- Bruce D Gunderson, Charles D Swerdlow, Jay M Wilcox, Jean E Hayman, Kevin T Ousdigian, and Kenneth A Ellenbogen.
- Medtronic, Inc., Minneapolis, Minnesota, USA.
- Heart Rhythm. 2010 May 1;7(5):626-33.
BackgroundImplantable cardioverter-defibrillator (ICD) ventricular oversensing may result in inappropriate therapy, which may be triggered by lead/connection issues that require surgical revision or physiologic oversensing that may be resolved with reprogramming. The sensing integrity counter (SIC) is an oversensing diagnostic that increments for very rapid ventricular intervals < or =130 ms.ObjectiveThe purpose of this study was to determine the causes of a high SIC and the ability of additional diagnostics to differentiate lead/connection issues from other causes of oversensing for patients with normal impedance.MethodsFrequent SICs were identified in patients during routine follow-up visits. To diagnose the cause of oversensing, patients wore a modified 24-hour digital Holter monitor that recorded ECG, ventricular electrogram, and the ICD Marker Channel (Medtronic). Recordings were reviewed to determine the causes of oversensing. Patients with confirmed oversensing and adequate data were analyzed. The number of SICs per day and the presence of a nonsustained tachycardia (NST) episode with ventricular mean cycle length <220 ms were retrieved from stored ICD data.ResultsForty-eight patients had a median of 13 SICs/day. Presumed lead/connection issues occurred in 23% of patients, whereas physiologic oversensing occurred in 77% of patients. A rapid NST was recorded more commonly in patients with lead/connection issues than in those without (9/11 vs 1/37; P < .0001).ConclusionOversensing resulting in frequent, very short intervals typically are caused by either lead/connection issues or physiologic signals. The additional finding of rapid NSTs usually indicates a lead/connection issue, even in the absence of impedance abnormalities.Copyright 2010 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.