Computerised lung sound monitoring to assess effectiveness of chest physiotherapy and secretion removal: a feasibility study.
- G Ntoumenopoulos and Y Glickman.
- Guy's and St Thomas' NHS Foundation Trust, London, UK. george.ntoumenopoulos@gstt.nhs.uk
- Physiotherapy. 2012 Sep 1;98(3):250-5.
ObjectivesTo explore the feasibility of computerised lung sound monitoring to evaluate secretion removal in intubated and mechanically ventilated adult patients.DesignBefore and after observational investigation.SettingIntensive care unit.ParticipantsFifteen intubated and mechanically ventilated adult patients receiving chest physiotherapy.InterventionsChest physiotherapy included combinations of standard closed airway suctioning, saline lavage, postural drainage, chest wall vibrations, manual-assisted cough and/or lung hyperinflation, dependent upon clinical indications.Main Outcome MeasuresLung sound amplitude at peak inspiration was assessed using computerised lung sound monitoring. Measurements were performed immediately before and after chest physiotherapy. Data are reported as mean [standard deviation (SD)], mean difference and 95% confidence intervals (CI). Significance testing was not performed due to the small sample size and the exploratory nature of the study.ResultsFifteen patients were included in the study [11 males, four females, mean age 65 (SD 14) years]. The mean total lung sound amplitude at peak inspiration decreased two-fold from 38 (SD 59) units before treatment to 17 (SD 19) units after treatment (mean difference 22, 95% CI of difference -3 to 46). The mean total lung sound amplitude from the lungs of patients with a large amount of secretions (n=9) was over four times 'louder' than the lungs of patients with a moderate or small amount of secretions (n=6) [56 (SD 72) units vs 12 (13) units, respectively; mean difference -44, 95% CI of difference -100 to 11]. The mean total lung sound amplitude decreased in the group of 'loud' right and left lungs (n=15) from 37 (SD 36) units before treatment to 15 (SD 13) units after treatment (mean difference 22, 95% CI of difference 6 to 38).ConclusionComputerised lung sound monitoring in this small group of patients demonstrated a two-fold decrease in lung sound amplitude following chest physiotherapy. Subgroup analysis also demonstrated decreasing trends in lung sound amplitude in the group of 'loud' lungs following chest physiotherapy. Due to the small sample size and large SDs with high variability in the lung sound amplitude measurements, significance testing was not reported. Further investigation is needed in a larger sample of patients with more accurate measurement of sputum wet weight in order to distinguish between secretion-related effects and changes due to other factors such as airflow rate and pattern.Copyright © 2012 Chartered Society of Physiotherapy. Published by Elsevier Ltd. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.