    An evaluation of the resistance to flow through the patient valves of twelve adult manual resuscitators.

    • D Hess and M Simmons.
    • York Hospital, PA 17405.
    • Resp Care. 1992 May 1;37(5):432-8.

    UnlabelledWhat is the inspiratory and expiratory resistance to flow through the patient valves of adult manual resuscitators?Materials & MethodsWe evaluated the resistance to flow through the patient valves of 12 adult resuscitators (Ambu, Code Blue, DMR, Hope 4, Hospitak, Hudson, Intertech, Laerdal, Mercury, Respironics, SPUR, Vitalograph). Expiratory resistance was evaluated by directing a flow of oxygen through the valve in the direction that the patient expires. Inspiratory resistance was evaluated by directing oxygen through the valve in the direction of flow when the bag is squeezed. Flow was controlled by a Timeter 0-75 flowmeter, and measured using a calibrated Timeter RT-200. Flows of 10, 20, 30, 40, 50, 60, 70, 80, and 90 L/min were used. Resistive back pressure of the resuscitator valves was measured using a calibrated Timeter RT-200. Resistance was calculated by dividing back pressure by flow. Five measurements were made at each flow setting for each resuscitator.ResultsSignificant differences in back pressures and resistances existed between the resuscitators for both expiratory and inspiratory flows (p less than 0.001 in each case). Significant interaction effects also existed between resuscitator brands and flows (p less than 0.001 in each case). At an expiratory flow of 50 L/min, all resuscitators except the Hospitak and Vitalograph produced a back pressure less than 5 cm H2O (the International Standards Organization standard). At an inspiratory flow of 50 L/min, all resuscitators but the Hospitak, Mercury, and Vitalograph produced a back pressure less than 5 cm H2O.ConclusionsSignificant differences existed in the back pressures produced due to the flow resistance through the patient valves of these resuscitators, and these might be considered excessive in some cases. Because this was a bench study, further work is needed to determine the clinical importance of these findings.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…