    Laboratory and clinical evaluation of the impact Uni-Vent 750 portable ventilator.

    • R S Campbell, K Davis, D J Johnson, D Porembka, and J M Hurst.
    • University of Cincinnati Medical Center, OH 45267.
    • Resp Care. 1992 Jan 1;37(1):29-36.

    BackgroundTransportation of critically ill, mechanically ventilated patients from intensive care units for diagnostic and therapeutic procedures has become common in the last decade. Maintenance of adequate oxygenation and ventilation during transport is essential. We evaluated the Impact Uni-Vent 750 portable ventilator in the laboratory and in the clinical arena to determine its usefulness during inhospital transport.Materials & MethodsIn the laboratory, we determined the Uni-Vent 750's ability to assure tidal volume (VT) delivery in the face of decreasing compliance of a test lung and tested the alarm systems. Using a two-compartment lung model modified to simulate spontaneous breathing, we also evaluated the responsiveness of the demand valve. The clinical evaluation was accomplished by comparing arterial blood gas values and ventilator settings in the intensive care unit before transport to those during transport.ResultsAs lung compliance was reduced from 0.1 to 0.02 mL/cm H2O [1.0 to 0.20 L/kPa], a slight, statistically insignificant decrease in delivered tidal volume was observed. All alarm systems operated according to manufacturer's specifications. The demand valve triggered appropriately with PEEP from 0 to 20 cm H2O [0 to 1.96 kPa]. Sensitivity settings less than -6 cm H2O [-0.59 kPa] sometimes resulted in inability to trigger the demand valve. During patient transport, arterial blood gas values and ventilator settings were comparable to those observed in the ICU. Because an FIO2 of 1.0 was used during transport, mean (SD) PaO2 was significantly greater 89 (26) vs 341 (78) [11.8 (3.5) vs 45.3 (10.4) kPa].ConclusionsThe Uni-Vent 750 is a reliable transport ventilator, capable of maintaining adequate oxygenation and ventilation in a majority of mechanically ventilated patients. The Uni-Vent 750's ability to (1) provide CMV, AMV, and SIMV; (2) provide low and high pressure alarms; and (3) provide PEEP compensation is unique among portable ventilators.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.