• Resp Care · Sep 2009

    Accuracy of the oxygen cylinder duration calculator of the LTV-1000 portable ventilator.

    • Thomas C Blakeman, Dario Rodriquez, and Richard D Branson.
    • Division of Trauma and Critical Care, Department of Surgery, University of Cincinnati Medical Center, 231 Albert Sabin Way, ML 0558, Cincinnati OH 45267-0558, USA. thomas.blakeman@uc.edu
    • Resp Care. 2009 Sep 1;54(9):1183-6.

    BackgroundResource planning is essential for successful transport of the mechanically ventilated patient. Mechanically ventilated patients require adequate oxygen supplies to ensure transport is completed without incident. The LTV-1000 portable ventilator utilizes a program to calculate oxygen cylinder duration, based on cylinder size, fraction of inspired oxygen (F(IO(2))), and current minute ventilation. We evaluated the accuracy of the cylinder-duration algorithm in a laboratory setting.MethodsThe LTV-1000 was attached to a test lung. Lung compliance was set at 0.04 L/cm H(2)O, and airway resistance was 5.0 cm H(2)O/L/s. We tested 7 different combinations of ventilator settings a minimum of 2 times each. With each setting, minute ventilation was kept at 10 L/min. Breath type, positive end-expiratory pressure, and F(IO(2)) were varied to evaluate the accuracy of the algorithm across a range of clinical scenarios. The cylinder-duration calculation from the ventilator program and manual calculation was determined at each setting and compared to the actual cylinder duration.ResultsThe ventilator algorithm and the manual calculation underestimated the actual cylinder duration by 12 +/- 3% with each test. The range of differences between calculated and actual cylinder duration was 2-26 min across the 7 conditions.ConclusionActual cylinder duration averaged 12% longer than the cylinder duration estimated by the algorithm of the LTV-1000. One explanation is that the E cylinders may contain more liters of oxygen than indicated by the sticker on the side of the tank. Additionally, the bias flow during expiration is affected by inspiratory-expiratory ratio and respiratory rate. Clinicians should be aware of these differences when planning for patient transport.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.