• Eur J Cardiothorac Surg · Mar 2010

    Rotation thromboelastometry (ROTEM) stability and reproducibility over time.

    • Oliver Michel Theusinger, Johannes Nürnberg, Lars M Asmis, Burkhardt Seifert, and Donat Rudolf Spahn.
    • Institute of Anesthesiology, University Hospital Zurich, CH - 8091 Zurich, Switzerland. oliver.theusinger@usz.ch
    • Eur J Cardiothorac Surg. 2010 Mar 1;37(3):677-83.

    BackgroundThromboelastometry is a whole blood assay performed to evaluate the viscoelastic properties during blood clot formation and lysis. Rotation thromboelastography (ROTEM), Pentapharm GmbH, Munich, Germany) has overcome some of the limitations of classic thromboelastography. So far, no clinical validation on reproducibility (inter- and intra-assay variability) and sample stability over time has been published.MethodsTo evaluate the pre-analytic aspects, sample stability over time was assessed in 48 patients in eight age groups. Citrated blood was stored at room temperature. Tests measured every 30 min from T 0 min up to T 120 min on two ROTEM devices were INTEM (ellagic acid activated intrinsic pathway), EXTEM (tissue factor-triggered extrinsic pathway) and FIBTEM (with platelet inhibitor (cytochalasin D) evaluating the contribution of fibrinogen to clot formation). Precision by intra- and inter-assay variability was evaluated at two points of time in 10 volunteers. Finally, reference intervals and effect of age and sex were evaluated.ResultsBlood was stable over 120 min and no significant differences in ROTEM results were found. Maximum clot firmness measurements had a coefficient of variation of <3% for EXTEM, <5% for INTEM and <6% for FIBTEM. For clot formation time, the coefficient of variation was <4% for EXTEM and <3% for INTEM. Coefficient of variation for angle alpha was <3% for EXTEM and <6% for INTEM. The coefficient of variation for clotting time was <15% for both EXTEM and INTEM. Small but significant differences between ROTEM devices were found for maximum clot firmness in FIBTEM and INTEM as well as clot formation time and alpha angle in INTEM.ConclusionsROTEM yields stable results over 120 min with a minimal variability on the same ROTEM device. However, small but significant differences between ROTEM devices were observed. Analysis should be performed on the same ROTEM device if small differences are of importance for treatment.Copyright (c) 2009 European Association for Cardio-Thoracic Surgery. Published by Elsevier B.V. All rights reserved.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…