Case Reports
Transcranial Doppler-guided deairing of a pediatric ventricular assist device: experience with twins.
- Gabor Erdoes, Alexander Kadner, Damian Hutter, and Balthasar Eberle.
- From the *Departments of Anesthesiology and Pain Therapy, †Cardiovascular Surgery, and ‡Cardiology, Inselspital, University Hospital Bern, Bern, Switzerland.
- A A Case Rep. 2015 Jan 1;4(1):5-7.
AbstractWe report the intraoperative courses of 2 consecutive Berlin Heart Excor® Pediatric Ventricular Assist Device implantations, in which transcranial Doppler ultrasonography helped to detect macroscopically undetected residual air bubbles captured in the pump after air removal had been correctly performed according to manufacturer's specifications. Our experience with these cases suggests that a beat-to beat deairing maneuver guided by transcranial Doppler is a useful strategy for reducing cerebral exposure to perioperative gaseous microembolism.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.