• Artificial organs · Jun 2000

    Closed circuit cardiopulmonary bypass with centrifugal pump for open-heart surgery: new trial for air removal.

    • M Morita, R Yozu, T Matayoshi, A Mitsumaru, H Shin, and S Kawada.
    • Medical Engineering Service, Keio University School of Medicine, Tokyo, Japan. morita@med.keio.ac.jp
    • Artif Organs. 2000 Jun 1;24(6):442-5.

    AbstractThe purpose of this study is to examine the efficiency of venous air removal with a new cardiopulmonary bypass (CPB) circuit design for conventional open-heart surgeries. A main concern with a closed circuit for open-heart surgeries is air entrainment into the venous line. A venous filter was placed proximal to the centrifugal pump. The circuit proximal to the centrifugal pump was divided into two lines; one line was attached to the venous reservoir outlet. By clamping the line to the reservoir, this circuit becomes closed. Negative pressure was applied to the purge line connected to the venous reservoir for venous air removal. Micro bubbles were measured at two locations, both distal to the venous and arterial filters. When the injection rate reached 100 ml/min, with the air-injection over 30 s, micro bubbles greater than 40 micro were observed at the outlet of venous filter. However, there was no micro bubble greater than 40 micro detected at the outlet of arterial filter. Although micro bubbles greater than 40 micro were not detected at the outlet of the arterial filter up to the injection rate of 300 ml/min, when the injection rate reached 400 ml/min, micro bubbles greater than 50 microm were detected distal to the arterial filter. From this examination, we determined that air entrained in the venous line up to approximately 300 ml/min is automatically removed by this method with the pressure-balanced condition. This pressure balance means that resistance of venous return, gravity siphon, negative pressure by centrifugal pump, and negative pressure applied to the air-purge line of the filter are balanced; that is, the venous return is sufficient, and the venous reservoir volume is kept stable. From this study we determined that this circuit design efficiently removes the entrained air in the venous line.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…