• Eur Spine J · Feb 2011

    The effect of design parameters of dynamic pedicle screw systems on kinematics and load bearing: an in vitro study.

    • C Schilling, S Krüger, T M Grupp, G N Duda, W Blömer, and A Rohlmann.
    • Research and Development, Biomechanical Research, Aesculap AG, Am Aesculap Platz, 78532, Tuttlingen, Germany. christoph.schilling@aesculap.de
    • Eur Spine J. 2011 Feb 1; 20 (2): 297-307.

    AbstractAs an alternative treatment for chronic back pain due to disc degeneration motion preserving techniques such as posterior dynamic stabilization (PDS) has been clinically introduced, with the intention to alter the load transfer and the kinematics at the affected level to delay degeneration. However, up to the present, it remains unclear when a PDS is clinically indicated and how the ideal PDS mechanism should be designed to achieve this goal. Therefore, the objective of this study was to compare different PDS devices against rigid fixation to investigate the biomechanical impact of PDS design on stabilization and load transfer in the treated and adjacent cranial segment. Six human lumbar spine specimens (L3-L5) were tested in a spine loading apparatus. In vitro flexibility testing was performed by applying pure bending moments of 7.5 Nm without and with additional preload of 400 N in the three principal motion planes. Four PDS devices, "DYN" (Dynesys(®), Zimmer GmbH, Switzerland), "DSS™" (Paradigm Spine, Wurmlingen, Germany), and two prototypes of dynamic rods, "LSC" with a leaf spring, and "STC" with a spring tube (Aesculap AG, Tuttlingen, Germany), were tested in comparison to a rigid fixation device S(4) (Aesculap AG, Tuttlingen, Germany) "RIG", to the native situation "NAT" and to a defect situation "DEF" of the specimens. The instrumented level was L4-L5. The tested PDS devices comprising a stiffness range for axial stiffness of 10 N/mm to 230 N/mm and for bending stiffness of 3 N/mm to 15 N/mm. Range of motion (ROM), neutral zone (NZ), and intradiscal pressure (IDP) were analyzed for all instrumentation steps and load cases of the instrumented and non-instrumented level. In flexion, extension, and lateral bending, all systems, except STC, showed a significant reduction of ROM and NZ compared to the native situation (p < 0.05). Furthermore, we found no significant difference between DYN and RIG (p > 0.1). In axial rotation, only DSS and STC reduced the ROM significantly (p < 0.005) compared to the native situation, whereas DYN and LSC stayed at the level of the native intersegmental rotation (p > 0.05). A correlation was found between axial stiffness and intersegmental stabilization in the sagittal and frontal plane, but not in the transversal plane where intersegmental stabilization is mainly governed by the systems' ability to withstand shear loads. Furthermore, we observed the systems' capacity to reduce IDP in the treated segment. The adjacent segment does not seem to be affected by the stiffness of the fixation device under the described loading conditions.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.