• Resp Care · Mar 1994

    Comparative Study

    A comparison of the CPAP performance characteristics of the Puritan-Bennett 7200a and a prototype continuous pressure-regulating ventilator.

    • R L Tobia and R J Fischer.
    • BOC Group Technical Center, Murray Hill, NJ 07974.
    • Resp Care. 1994 Mar 1; 39 (3): 195-203.

    BackgroundA prototype demand-flow medical ventilator for intensive care unit (ICU) applications has been developed with the ability to maintain continuous pressure regulation of proximal airway pressure during both inspiratory and expiratory respiratory phases. The performance of this system was investigated in laboratory tests of continuous positive airway pressure (CPAP) mode, a ventilatory mode in which airway pressure regulation is strongly challenged.Materials & MethodsComparative tests of the pressure-regulating ventilator (PRV) prototype and a Puritan-Bennett 7200a (PB7200a) ventilator were made in three performance categories: pressure-volume product error, peak pressure error during inspiration, and peak pressure error during expiration. Testing was performed on a Bio-Tek VT-2 test lung modified to simulate spontaneous breathing, at CPAP levels of 0, 10, and 20 cm H2O. Results were obtained at a lung compliance of 50 mL/cm H2O and at three parabolic airway resistance levels (Rp10, Rp20, and Rp50 cm H2O.L-2.s2 per ANSI Z79.7-1976).ResultsThe PRV prototype achieved reductions in pressure-volume product error of more than 0.111 joules (43%) for all test conditions, relative to the PB7200a. Peak pressure errors were reduced by at least 4.0 cm H2O (55%) during inspiration and 4.9 cm H2O (38%) during expiration.ConclusionsAn ICU ventilator utilizing continuous regulation of proximal airway pressure has been shown to provide statistically significant improvements in CPAP performance relative to the PB7200a in laboratory tests. Further testing will be required to quantify the clinical significance of these results.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.