• Acta Neurochir. Suppl. · Jan 1998

    Comparative Study Clinical Trial

    Dual-switch valve: clinical performance of a new hydrocephalus valve.

    • H A Trost, C Sprung, W Lanksch, D Stolke, and C Miethke.
    • Neurosurgical Department, Krankenhaus Bogenhausen, Munich, Federal Republic of Germany.
    • Acta Neurochir. Suppl. 1998 Jan 1; 71: 360-3.

    AbstractThe Dual-Switch valve (DSV) is the first construction on the market which changes between two different valve-chambers in parallel depending on the posture of the patient. In the lying position the valve acts like a conventional differential pressure valve, in the vertical position the high-pressure chamber only opens, when the pressure exceeds the hydrostatic pressure difference between the formanen of Monro and the peritoneal cavity. The new device has been implanted in 32 adult patients with hydrocephalus of different etiology. The clinical results are excellent to good accompanied by a remarkable slight reduction of the ventricular size. Apart from one case with a nonsymptomatic transient hygroma, we saw no valve related complications like overdrainage, underdrainage or dysfunction. Contrary to conventional differential-pressure valves, adjustable devices and other hydrostatic constructions like the Anti-Siphon-device (ASD) or Deltavalve, the DSV reliably controls the IVP independently of the posture of the patient, the CSF viscosity or the subcutaneous pressure. In contrast to the Orbis-Sigma-valve (OSV) or the Diamond-valve, the DSV does not control the flow but the physiological IVP avoiding the increased risk of mechanical failure. The results of this study give strong evidence that the shunt-therapy of adult hydrocephalic patients can be significantly improved by the DSV.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…