    Three-dimensional rotational angiography of the inferior vena cava as an adjunct to inferior vena cava filter retrieval.

    • Ugur Bozlar, J Stewart Edmunds, Ulku C Turba, Gary D Hartwell, Ahmed M Housseini, and Klaus D Hagspiel.
    • Department of Radiology, University of Virginia Health System, Charlottesville, VA 22908, USA.
    • Cardiovasc Intervent Radiol. 2009 Jan 1; 32 (1): 86-92.

    AbstractThe objective of this study was to explore the role of three-dimensional (3-D) rotational angiography (RA) of the inferior vena cava (IVC; 3-D CV) before filter retrieval and its impact on treatment planning compared with standard anteroposterior cavography (sCV). Thirty patients underwent sCV and 3-D CV before IVC filter retrieval. Parameters assessed were: projection of filter arms or legs beyond the caval lumen, thrombus burden within the filter and IVC, and orientation of the filter within IVC. Skin and effective radiation doses were calculated. Statistical analysis was performed using paired Student t test and nonparametric McNemar's test. Standard anteroposterior cavography detected 49 filter arms or legs projecting beyond the caval lumen in 25 patients. Three-dimensional CV demonstrated 89 filter arms or legs projecting beyond the caval lumen in 28 patients. Twenty-two patients had additional filter arms or legs projecting beyond the caval lumen detected on 3-D CV that were not detected on sCV (p < 0.001). Filter apex tilt detection differed significantly (p < 0.001) between sCV and 3-D CV, with 3-D CV being more accurate. The filter apex abutted the IVC wall in 10 patients (33%) on 3-D CV, but this was diagnosed in only 3 patients (10%) with sCV. Thrombus was detected in 8 patients (27%), 1 thrombus of which was seen only on 3-D CV, and treatment was changed in this patient because of thrombus size. Mean effective radiation doses for 3-D CV were approximately two times higher than for sCV (1.68 vs. 0.86 mSv), whereas skin doses were three times lower (12.87 vs. 35.86 mGy). Compared with sCV, performing 3-D CV before optional IVC filter retrieval has the potential to improve assessment of filter arms or legs projecting beyond the caval lumen, filter orientation, and thrombus burden.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…