• Respiratory care · Jul 2018

    Observational Study

    Accuracy of Transcutaneous CO2 Values Compared With Arterial and Capillary Blood Gases.

    • Laura L Lambert, Melissa B Baldwin, Cruz Velasco Gonzalez, Gary R Lowe, and J Randy Willis.
    • Respiratory Care Services, Arkansas Children's Hospital, Little Rock, Arkansas.
    • Respir Care. 2018 Jul 1; 63 (7): 907-912.

    BackgroundTranscutaneous monitors are utilized to monitor a patient's respiratory status. Some patients have similar values when comparing transcutaneous carbon dioxide (PtcCO2 ) values with blood gas analysis, whereas others show extreme variability. A retrospective review of data was performed to determine how accurately PtcCO2 correlated with CO2 values obtained by arterial blood gas (ABG) or capillary blood gas.MethodsTo determine whether PtcCO2 values correlated with ABG or capillary blood gas values, subjects' records were retrospectively reviewed. Data collected included the PtcCO2 value at the time of blood gas procurement and the ABG or capillary blood gas PCO2 value. Agreement of pairs of methods (ABG vs PtcCO2 and capillary blood gas vs PtcCO2 ) was assessed with the Bland-Altman approach with limits of agreement estimated with a mixed model to account for serial measurements per subject.ResultsA total of 912 pairs of ABG/PtcCO2 values on 54 subjects and 307 pairs of capillary blood gas/PtcCO2 values on 34 subjects were analyzed. The PCO2 range for ABG was 24-106 mm Hg, and PtcCO2 values were 27-133 mm Hg. The PCO2 range for capillary blood gas was 29-108 mm Hg, and PtcCO2 values were 30-103 mm Hg. For ABG/PtcCO2 comparisons, the Pearson correlation coefficient was 0.82, 95% CI was 0.80-0.84, and P was <.001. For capillary blood gas/PtcCO2 comparisons, the Pearson correlation coefficient was 0.77, 95% CI was 0.72-0.81, and P was <.001. For ABG/PtcCO2 , the estimated difference ± SD was -6.79 ± 7.62 mm Hg, and limits of agreement were -22.03 to 8.45. For capillary blood gas/PtcCO2 , the estimated difference ± SD was -1.61 ± 7.64 mm Hg, and limits of agreement were -16.88 to 13.66. The repeatability coefficient was about 30 mm Hg.ConclusionsBased on these data, capillary blood gas comparisons showed less variation and a slightly lower correlation with PtcCO2 than did ABG comparisons. After accounting for serial measurements per patient, due to the wide limits of agreement and poor repeatability, the utility of relying on PtcCO2 readings for this purpose is questionable.Copyright © 2018 by Daedalus Enterprises.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.