• Respiratory care · Jul 2018

    Comparative Study

    A Comparison of the Adjustable Ranges of Inspiratory Pressurization During Pressure Controlled Continuous Mandatory Ventilation of 5 ICU Ventilators.

    • Yukiko Koyama, Akinori Uchiyama, Junko Yoshida, Takeshi Yoshida, Tomonori Yamashita, and Yuji Fujino.
    • Department of Anesthesiology and Intensive Care Medicine, Osaka University Graduate School of Medicine.
    • Respir Care. 2018 Jul 1; 63 (7): 849-858.

    BackgroundFaster inspiratory pressurization can improve patient-ventilator synchrony and reduce the patient's work of breathing during pressure controlled continuous mandatory (PC-CMV) ventilation. The characteristics of the pressurization ramp settings are not standardized across ventilators from different manufacturers. We performed a bench test of 5 models of ICU ventilators to examine the effects of pressurization ramp settings on the actual pressurization.MethodsA twin-bellows lung model was used, in which one bellow simulates inspiratory muscle activity and the other simulates an adult normal lung model. We made the inspiratory effort by changing the tidal volume of the inspiratory muscle bellow. The effect of pressurization ramp settings on the performance of each ventilator was examined at 3 inspiratory effort levels (ie, none, ordinary, and strong). The pressurization ramp was set at 4 or 5 evenly divided steps from the minimum to maximum for each ventilator. The following parameters were measured: tidal volume, mean airway pressure, maximal inspiratory flow, time to maximal flow, and pressure-time products at 0.3 s (PTP0.3) and 0.5 s (PTP0.5) from the beginning of inspiration. PTP0.3 and PTP0.5 indicated levels of inspiratory pressurization.ResultsA proportional increase in PTP0.3 and PTP0.5 was observed with an increase in the pressurization ramp settings of the recent models of ventilators. PTP0.3 and PTP0.5 at ordinary and strong effort levels were similar in the recent models of ventilators. The actual adjustable ranges of PTP0.3 and PTP0.5 associated with change in the pressurization ramp settings differed between the 5 ventilators.ConclusionsThe adjustable ranges of the pressurization were largely different among the different types of ventilators. The actual absolute inspiratory pressurization during PC-CMV varied between the different ventilators even at similar pressurization ramp settings. Users should be mindful of the differences in the pressurization ramp settings.Copyright © 2018 by Daedalus Enterprises.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…