• Artificial organs · Jan 2019

    Hemodynamic Evaluation of Avalon Elite Bi-Caval Dual Lumen Cannulas and Femoral Arterial Cannulas.

    • Shigang Wang, Madison Force, Allen R Kunselman, David Palanzo, Christoph Brehm, and Akif Ündar.
    • Department of Pediatrics, Penn State Health Pediatric Cardiovascular Research Center, Penn State Milton S. Hershey Medical Center, Penn State College of Medicine, Penn State Health Children's Hospital, Hershey, PA, USA.
    • Artif Organs. 2019 Jan 1; 43 (1): 41-53.

    AbstractTranslational research is a useful tool to provide scientific evidence for cannula selection during extracorporeal life support (ECLS). The objective of this study was to evaluate four Avalon Elite bi-caval dual lumen cannulas and nine femoral arterial cannulas in terms of flow range, circuit pressure, pressure drop, and hemodynamic energy transmission in a simulated adult ECLS model. A veno-venous ECLS circuit was used to evaluate four Avalon Elite bi-caval dual lumen cannulas (20, 23, 27, and 31 Fr), and a veno-arterial ECLS circuit was used to evaluate nine femoral arterial cannulas (15, 17, 19, 21, and 23 Fr). The two circuits included a Rotaflow centrifugal pump, a Quadrox-D adult oxygenator, and 3/8 in ID tubing for arterial and venous lines. The circuits were primed with lactated Ringer's solution and packed human red blood cells (hematocrit 40%). Trials were conducted at rotational speeds from 1000 to 5000 RPM (250 rpm increments) for each Avalon cannula, and at different flow rates (0.5-7 L/min) for each femoral arterial cannula. Real-time pressure and flow data were recorded for analysis. Small caliber cannulas created higher circuit pressures, higher pressure drops and higher M-numbers compared with large ones. The inflow side of Avalon dual lumen cannula had a significantly higher pressure drop than the outflow side (inflow vs. outflow: 20 Fr-100.2 vs. 49.2 mm Hg at 1.1 L/min, 23 Fr-93.7 vs. 41.4 mm Hg at 1.6 L/min, 27 Fr-102.3 vs. 42.8 mm Hg at 2.6 L/min, 31 Fr-98.1 vs. 44.7 mm Hg at 3.8 L/min). There was more hemodynamic energy lost in the veno-arterial ECLS circuit using small cannulas compared to larger ones (17 Fr vs. 19 Fr vs. 21 Fr at 4 L/min-Medtronic: 71.0 vs. 64.8 vs. 60.9%; Maquet: 71.4 vs. 65.6 vs. 62.0%). Medtronic femoral arterial cannulas had lower pressure drops (Medtronic vs. Maquet at 4 L/min: 17 Fr-121.7 vs. 125.0 mm Hg, 19 Fr-71.2 vs. 73.7 mm Hg, 21 Fr-42.9 vs. 47.4 mm Hg) and hemodynamic energy losses (Medtronic vs. Maquet at 4 L/min: 17 Fr-43.6 vs. 44.4%, 19 Fr-31.0 vs. 31.4%, 21 Fr-20.8 vs. 22.4%) at high flow rates when compared with the Maquet cannulae. The results for this study provided valuable hemodynamic characteristics of all evaluated adult cannulas with human blood in order to guide ECLS cannula selection in clinical practice. Use of larger cannulas are suggested for VV- and VA-ECLS.© 2018 International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…