-
J Clin Monit Comput · Dec 1999
Inhaled nitric oxide fraction is influenced by both the site and the mode of administration.
- E Sieffert, L Ducros, M R Losser, and D M Payen.
- Département d'Anesthésie-Réanimation & SMUR, Assistance Publique-Hôpitaux de Paris, Hôpital Lariboisière, 2 rue Ambroise Paré, 75475 Paris Cedex 10, France.
- J Clin Monit Comput. 1999 Dec 1; 15 (7-8): 509-17.
ObjectiveInhaled nitric oxide (NO) can be delivered continuously or sequentially (= during inspiration) at different locations of the ventilation circuit. We have tested the influence of locations, modes of NO administration and the ratio of the inspiratory time over the respiratory cycle time (I/I + E ratio) on the accuracy of NO fractions, delivered by 2 devices: Opti-NO and Flowmeter.MethodsWe used a simplified lung model consisting of a ventilation circuit with a Y piece, a tracheal tube, a 150 ml dead-space volume and a 5 liter balloon. Three fractions (3, 6, 9 ppm) were administered continuously or sequentially, in controlled volume, in 4 different sites on the inspiratory branch above the Y piece: i) just after the water trap, ii) just before the Y piece; below the Y piece: iii) just after the Y piece, iv) into the endotracheal tube. In addition, different I/I + E ratios (25, 33, 50, 80%) were studied. The delivered NO fractions were measured in the balloon by chemiluminescence (CLD 700, Ecophysics). A linear regression analysis was used to test the relationship between administered and measured NO fractions for the 3 fractions (3, 6 and 9 ppm) in sequential and continuous modes. Intercept values were compared to zero and slopes to the identity line.ResultsWhen NO was administered in the continuous mode upstream the Y piece, NO fractions measured in the balloon corresponded to the administered fractions. In contrast, below the Y piece, the measured NO fractions were significantly lower than the administered NO fractions. In the sequential mode, above and below the Y piece, the delivered NO fractions were within the manufacturer's range.ConclusionsFor the continuous NO delivery, locations above the Y piece are mandatory. However, locations below the Y piece imposes a sequential system, which can also be used for the sites located above the Y piece.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.