Comparative StudyDo Epidural Catheter Size and Flow Rate Affect Bolus Injection Pressure in Different Programmed Intermittent Epidural Bolus Regimens? An In Vitro Study.
- Paweł Krawczyk, Piotr Piwowar, Kinga Sałapa, Tomasz Lonc, and Janusz Andres.
- From the Department of Anesthesiology and Intensive Care Medicine, Jagiellonian University Medical College, Cracow, Poland.
- Anesth. Analg. 2019 Dec 1; 129 (6): 158715941587-1594.
BackgroundThe optimal programmed intermittent epidural bolus regimen for labor analgesia remains unknown. Some studies indicate that better drug spread in the epidural space results from greater injection pressure; however, there is a lack of data regarding the maximum pressure generated by epidural bolus injection using different catheters and flow rates.MethodsWe evaluated the flow and pressure characteristics of 11 commonly used epidural catheters combined with 3 different infusion pumps that deliver epidural infusions according to the programmed intermittent epidural bolus regimen. Pressure changes were measured over time at flow rates of 100, 250, and 400 mL·hour and with a bolus volume of 10 mL. To account for repeated measures, linear mixed models were used. Features were selected with a backward stepwise procedure continued until only statistically significant variables were left in the model.ResultsWe performed 660 measurements. The mean maximal pressure generated during bolus injection ranged from 86 to 863 mm Hg for different flow rates and catheter designs. The interaction between flow rate and catheter gauge resulted in 1.31, 1.65, and 2.00 mm Hg of pressure increase for 18G, 19G, and 20G catheters, respectively, per 1 mL·hour of increased flow rate (P< .001). Analyses including wire-reinforced catheters revealed a 1.16, 1.76, and 2.36 mm Hg pressure increase for 18G, 19G, and 20G catheters, respectively, per 1 mL·hour of increased flow rate (P< .001). In some cases, it triggered the occlusion pump alarm.ConclusionsSignificant differences were observed in the in vitro maximum pressure value among the various catheter and flow rate combinations with a higher pressure value for wire-reinforced catheters used in the study. The optimal flow rate and epidural catheter combination may allow for delivery of the bolus with high flow rate without triggering the occlusion alarm.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.