- Shigang Wang, Jenelle M Izer, Joseph B Clark, Sunil Patel, Linda Pauliks, Allen R Kunselman, Donald Leach, Timothy K Cooper, Ronald P Wilson, and Akif Ündar.
- Department of Pediatrics, Penn State Hershey Pediatric Cardiovascular Research Center, Penn State Milton S. Hershey Medical Center, Penn State Hershey College of Medicine, Penn State Hershey Children's Hospital, Hershey, PA, USA.
- Artif Organs. 2015 Jul 1; 39 (7): E90-E101.
AbstractThe primary objective of this study was to evaluate a novel electrocardiogram (ECG)-synchronized pulsatile extracorporeal life support (ECLS) system for adult partial mechanical circulatory support for adequate quality of pulsatility and enhanced hemodynamic energy generation in an in vivo animal model. The secondary aim was to assess end-organ protection during nonpulsatile versus synchronized pulsatile flow mode. Ten adult swine were randomly divided into a nonpulsatile group (NP, n = 5) and pulsatile group (P, n = 5), and placed on ECLS for 24 h using an i-cor system consisting of an i-cor diagonal pump, an iLA membrane ventilator, an 18 Fr femoral arterial cannula and a 23/25 Fr femoral venous cannula. Trials were conducted at a flow rate of 2.5 L/min using nonpulsatile or pulsatile mode (with assist ratio 1:1). Real-time pressure and flow data were recorded using a custom-based data acquisition system. To the best of our knowledge, the oxygenator and circuit pressure drops were the lowest for any available system in both groups. The ECG-synchronized i-cor ECLS system was able to trigger pulsatile flow in the porcine model. After 24-h ECLS, energy equivalent pressure, surplus hemodynamic energy, and total hemodynamic energy at preoxygenator and prearterial cannula sites were significantly higher in the P group than those in the NP group (P < 0.05). Urine output was higher in P versus NP (3379 ± 443 mL vs. NP, 2598 ± 1012 mL), and the P group seemed to require less inotropic support, but both did not reach statistical significances (P > 0.05). The novel i-cor system performed well in the nonpulsatile and ECG-synchronized pulsatile mode in an adult animal ECLS model. The iLA membrane oxygenator had an extremely lower transmembrane pressure gradient and excellent gas exchange capability. Our findings suggest that ECG-triggered pulsatile ECLS provides superior end-organ protection with improved renal function and systemic vascular tone. Copyright © 2015 International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.