A combination of local flow measurement techniques and fog flow visualization was used to determine the inward leakage for two tight-fitting powered air-purifying respirators (PAPRs), the 3M Breathe-Easy PAPR and the SE 400 breathing demand PAPR. The PAPRs were mounted on a breathing machine head form, and flows were measured from the blower and into the breathing machine. Both respirators leaked a little at the beginning of inhalation, probably through their exhalation valves. In both cases, the leakage was not enough for fog to appear at the mouth of the head form.
-
J Environ Public Health · Jan 2011
Comparative StudyUsing CO(2) to determine inhaled contaminant volumes and blower effectiveness in several types of respirators.
This experiment was conducted to determine how much contaminant could be expected to be inhaled when overbreathing several different types of respirators. These included several tight-fitting and loose-fitting powered air-purifying respirators (PAPRs) and one air-purifying respirator (APR). CO(2) was used as a tracer gas in the ambient air, and several loose-and tight-fitting respirators were tested on the head form of a breathing machine. ⋯ Blower effectiveness was about 1.0 for tight-fitting APRs, 0.18 for the Racal, and greater than 1.0 for two of the loose-fitting PAPRs. With blower effectiveness greater than 1.0, some blower flow during the exhalation phase contributes to the subsequent inhalation. Results from this experiment point to different ways to measure respirator efficacy.