ASAIO journal : a peer-reviewed journal of the American Society for Artificial Internal Organs
-
To balance the risk of bleeding versus circuit thrombosis during extracorporeal life support (ECLS), it is important to monitor anticoagulants and hemostasis. We evaluated the prothrombin time (PT), partial thromboplastin time (PTT), activated clotting time (ACT), and antifactor Xa heparin activity (aXa) correlation with changes in coagulation factor and heparin levels using in vitro and in vivo samples. aXa correlated with heparin (r = 0.97) and antithrombin (r = 0.98) but was unaffected by other parameters. PT correlated with coagulation factors (r = 0.88) but was minimally affected by heparin or other parameters. ⋯ In conclusion, aXa is the most specific for heparin levels, and PT is most specific for coagulation factor levels making these assays well suited to monitor anticoagulation and hemostasis for patients on ECLS. PTT is highly variable when multiple parameters are changing in vitro and in vivo, but may be useful when aXa cannot be used because of interference. ACT is too insensitive to heparin, sensitive to too many other factors, and too imprecise to be useful for monitoring hemostasis during ECLS.