Technology and health care : official journal of the European Society for Engineering and Medicine
-
Keeping the oxygenation status of newborn infants within physiologic limits is a crucial task in intensive care. For this purpose several vital parameters are supervised routinely by monitors, such as electrocardiograph, transcutaneous partial oxygen pressure monitor and pulse oximeter. Each monitor issues an alarm signal whenever an upper or lower limit of the parameter(s) measured is exceeded. ⋯ The data and video were analysed off line with the help of an experienced neonatologist. His judgement was compared to the analysis of the Fuzzy Logic system. The results show that it is possible to improve the reliability of the monitored data with the aid of an evaluation strategy based on Fuzzy Logic and hence distinguish between real alarm situations and movement artefacts to the extent that an application in an intensive care unit under routine conditions becomes conceivable.