At high ambient temperatures in ICU rooms, the humidification performances of heated-wire humidifiers are significantly reduced, with delivered gas humidity well below 30 mg H2O/L, which leads to an increased risk of endotracheal occlusions, subocclusions, or mucociliary dysfunction. The objective of the study was to evaluate the humidity delivered at the Y-piece with new-generation heated-wire humidifiers with advanced algorithm (FP950 [Fisher & Paykel Healthcare, Auckland, New Zealand] and VHB20 [Vincent Medical, Inspired, Hong Kong]) while varying ambient temperatures. ⋯ The new FP950 and VHB20 heated-wire humidifiers by using advanced algorithms demonstrated stable performance while varying the ambient temperature by 20-30°C, better than the previous generation of heated humidifiers when ambient temperatures were high.
-
In children with congenital heart disease, extubation readiness testing (ERT) is performed to evaluate the potential for liberation from mechanical ventilation. There is a paucity of data that suggests what mechanical ventilation parameters are associated with successful ERT. We hypothesized that ERT success would be associated with certain mechanical ventilator parameters. ⋯ ERT pass rates decreased as ventilator support increased; however, some subjects were able to pass ERT despite high ventilator support. We found that [Formula: see text] was associated with ERT success and that protocols should consider using [Formula: see text] instead of PEEP thresholds for ERT eligibility. Cyanotic lesions were not associated with ERT success, which suggests that patients with cyanotic heart disease can be included in ERT protocols.