New equipment to prevent carbon dioxide rebreathing during eye surgery under retrobulbar anaesthesia.
Carbon dioxide concentration under ophthalmic drapes increases during eye surgery under local anaesthesia. A new prototype has been designed which combines continuous suction of carbon dioxide enriched air and continuous oxygen insufflation under ophthalmic drapes to prevent carbon dioxide accumulation in spontaneously breathing patients undergoing cataract surgery. ⋯ Application of this new prototype for continuous aspiration of carbon dioxide enriched air prevents carbon dioxide rebreathing and subsequent hypercapnia associated with an elevated respiratory rate. This new equipment may therefore be useful in patients undergoing ophthalmic surgery under retrobulbar anaesthesia.