Comparative StudyContinuous cardiac output monitoring with an uncalibrated pulse contour method in patients supported with mechanical pulsatile assist device.
We evaluated the accuracy of an uncalibrated pulse contour method called Pressure Recording Analytical Method (PRAM) compared with continuous thermodilution for cardiac output (CO) monitoring in patients implanted with a pulsatile left ventricular assist device (LVAD). Twelve adult patients implanted with the HeartMate I-XVE device were studied. CO was simultaneously evaluated by PRAM and by continuous thermodilution. ⋯ A good correlation was found between LVAD-CO and either ThD-CCO (r=0.88) or PRAM-CO (r=0.86), but an overestimation of 10% was observed for both PRAM-CO (mean bias -0.44 l/min) and ThD-CCO (mean bias -0.40 l/min). Our results demonstrated good agreements between PRAM-CO, ThD-CCO and LVAD-CO. PRAM derives CO from a peripheral artery without calibration and may be a complementary tool in the hemodynamic assessment of patients supported with a VAD.