Chest
-
To minimize work of breathing, airway pressure should not fluctuate during spontaneous breathing with continuous positive airway pressure (CPAP). However, flow resistance in the inspiratory limb of the breathing circuit and an inadequate continuous gas flow rate result in airway pressure fluctuation and increased work of breathing. ⋯ Work increased up to threefold with the balloon valve and more than tenfold with the flow resistor compared with the threshold resistor. To apply CPAP, expiratory pressure valves with low resistance to flow should be used to minimize fluctuations in airway pressure and, thus, in the work of spontaneous breathing.