Introduction of a Novel System for Quantitating Blood Loss After Vaginal Delivery: A Retrospective Interrupted Time Series Analysis With Concurrent Control Group.
Postpartum hemorrhage (PPH) is a leading cause of maternal mortality worldwide. While volumetric and gravimetric blood loss measurement techniques have been shown to increase PPH detection compared to visual estimation of blood loss (vBL), a combination of these methods has not been evaluated. The primary aim of this study was to compare the change in odds of PPH detection after vaginal delivery (VD; the intervention series) versus cesarean delivery (CD; the control series) before versus after implementation of a device (Triton L&D, Gauss Surgical, Inc, Los Altos, CA) that combines gravimetric and volumetric estimation of blood loss (tBL). ⋯ In this interrupted time series analysis, the odds of PPH detection after VD increased after implementation of the tBL device. However, a statistical difference was not detected between the increase in PPH detection post- versus preimplementation of the device in VDs versus CDs.