Effects of different triggering systems and external PEEP on trigger capability of the ventilator.
The triggering capability of both the pressure and flow triggering systems of the Servo 300 ventilator (Siemens-Elema, Sweden) was compared at various levels of positive end-expiratory pressure (PEEP), airway resistance (R(aw)), inspiratory effort and air leak, using a mechanical lung model. ⋯ With respect to triggering delay, the triggering capabilities of the pressure and flow sensing systems were comparable with and without PEEP and/or high airway resistance at the same sensitivity level, unless low inspiratory drive and air leak were present. In terms of pressure differentials, the flow triggering system may require less inspiratory effort to trigger the ventilator than that of the pressure triggering system with a comparable triggering time. However, this difference may be extremely small.