Resuscitation
-
Revision open heart surgery may be impeded by a dense network of pericardial adhesions rendering cardiac mobilization laborious or incomplete, and internal defibrillation impossible. External defibrillation, the current alternative to internal defibrillation, may result in myocardial stunning secondary to the delivery of escalating, monophasic, high-energy shocks. Automated external defibrillation, by delivering consecutive, non-escalating, impedance-compensated, low-energy, biphasic electric shocks to the myocardium, may provide a more effective and safer option whilst reducing the risk of myocardial stunning.
-
This study examined (1) the defibrillation efficacy of using a minimally invasive direct cardiac massage (MID-CM) device as one electrode of the defibrillation electrical circuit and (2) the effect on external defibrillation of defibrillating when the MID-CM device is in place and a pneumothorax is present. ⋯ Using the MID-CM device as one electrode of the defibrillation circuit markedly lowers the DFT compared with that for standard transthoracic defibrillation for both a monophasic and biphasic waveform. Defibrillation with the device in place and the chest opened elevates the DFT for external defibrillation much more during end-expiration than during end-inspiration.