Comparison of myelin water fraction from multiecho T2 decay curve and steady-state methods.
Myelin water fraction is conventionally measured from the T2 decay curve. Recently, a steady-state approach entitled multicomponent-driven equilibrium single pulse observation of T1 /T2 (mcDESPOT) was employed for myelin water fraction mapping. However, no direct comparison between the established multiecho T2 relaxation method and mcDESPOT has been performed. ⋯ Myelin water fraction values derived from mcDESPOT cannot be considered to be equivalent to those derived from T2 decay curve approaches.
-
To develop a three-dimensional (3D) free-breathing myocardial T1 mapping sequence for assessment of left ventricle diffuse fibrosis after contrast administration. ⋯ The proposed sequence enables high-resolution 3D T1 mapping after contrast injection during free-breathing with volumetric left ventricle coverage.
-
To propose a new phase-based B1-mapping method that exploits phase information created by hyperbolic secant (HS) pulses in conventional 2D spin-echo imaging. ⋯ The proposed method is easy to implement without any sequence modification, is insensitive to B0 inhomogeneity and chemical shift, and is robust in a reasonably wide range of B1 field strength.
-
Frequency and phase drifts are a common problem in the acquisition of in vivo magnetic resonance spectroscopy (MRS) data. If not accounted for, frequency and phase drifts will result in artifactual broadening of spectral peaks, distortion of spectral lineshapes, and a reduction in signal-to-noise ratio (SNR). We present herein a new method for estimating and correcting frequency and phase drifts in in vivo MRS data. ⋯ Spectral registration provides an effective method for frequency and phase drift correction. It does not involve the collection of navigator echoes, and does not rely on any specific resonances, such as residual water or creatine, making it highly versatile.
-
Comparative Study
Quantitative cerebral blood flow with bolus tracking perfusion MRI: measurements in porcine model and comparison with H(2)(15)O PET.
We recently presented a method for the quantitative measurement of the arterial input function which allows for determination of absolute cerebral blood flow (CBF) values without adjustable parameters. The aim of the present work is to estimate absolute CBF values by using this new technique and to compare it with the gold standard for cerebral perfusion, H(2)(15)O positron emission tomography. ⋯ The proposed method allows for determination of quantitative CBF without normalization factors. The relatively low estimates of absolute CBF most likely results from the higher age of the pigs as compared to other studies. The intermediate correlation between both methods is caused by physiological intraindividual fluctuations of the CBF and by a limited reproducibility of both methods.