Perfusion
-
We describe an adverse event during minimally invasive cardiac surgery that resulted in a multi-disciplinary review of intra-operative errors and the creation of a procedural checklist. This checklist aims to prevent errors of omission and communication failures that result in increased morbidity and mortality. We discuss the application of the aviation - led "threats and errors model" to medical practice and the role of checklists and other strategies aimed at reducing medical errors.
-
Gaseous microemboli (GME) generated during cardiopulmonary bypass (CPB) can present a significant risk to patient outcomes, specifically if they are delivered to the cerebral vasculature. A number of GME sources have been identified, leading to improved clinical practice and equipment design to ameliorate the presence and intensity of GME during CPB. Recently, a number of new venous reservoir/oxygenator systems have entered the market, including the Sorin Inspire6 and Inspire8, the Terumo FX15 and FX25 and the Maquet Quadrox-i. ⋯ The FX15 and FX25 were least effective removing small GME, while the FX15 and Quadrox-i were the least effective at removing medium GME. The Quadrox-i was least effective at removing large GME. In terms of complete venous reservoir/oxygenator systems, the Synthesis permitted the greatest amount of GME to pass, while the other systems appeared largely equivalent.