In an attempt to preserve proximal femoral bone stock and achieve a better fit in smaller femora, especially in the Asian population, several new shorter stem designs have become available. We investigated the torque to periprosthetic femoral fracture of the Exeter short stem compared with the conventional length Exeter stem in a Sawbone model. ⋯ Results showed that Sawbone femurs break at a statistically significantly lower torque to failure with a shorter compared to conventional-length Exeter stem of the same offset. Both standard and short-stem designs are safe to use as the torque to failure is 7-10 times that seen in activities of daily living.
-
Bone ingrowth into uncemented acetabular components requires intimate cup-bone contact and rigid fixation, which can be difficult to achieve in revision hip arthroplasty. This study compares polyaxial compression locking screws with non-locked and cancellous screw constructs for acetabular cup fixation. An acetabular cup modified with screw holes to provide both compression and angular stability was implanted into a bone substitute. ⋯ Polyaxial locking compression screws significantly improved construct stiffness compared with non-locked or cancellous screws. This increased construct stiffness will likely reduce interfacial micromotion. Further research is required to determine whether this will improve bone ingrowth in vivo and reduce cup failure.