Surgical endoscopy
-
Electromagnetic coupling can occur between the monopolar "Bovie" instrument and other laparoscopic instruments without direct contact by a phenomenon termed antenna coupling. The purpose of this study was to determine if, and to what extent, radiofrequency energy couples to other common laparoscopic instruments and to describe practical steps that can minimize the magnitude of antenna coupling. ⋯ Commonly used laparoscopic instruments couple monopolar radiofrequency energy without direct contact with the active electrode, a phenomenon that results in heat transfer from a nonelectrically active instrument tip to adjacent tissue. Practical steps to minimize heat transfer resulting from antenna coupling include reducing the monopolar generator power setting and avoiding of parallel bundling of the telescope and active electrode cords.