Anaesthesia
-
We studied the effect of pilot balloon design on the ability of experienced anaesthetists to assess and inflate tracheal tube cuffs to safe pressures. A model trachea was designed, incorporating a degree of compliance and an air leak, to evaluate six different pilot balloons grafted onto identical tracheal tubes. Pilot balloons were inflated to one of four pressures and anaesthetists were asked to estimate whether the pressure was acceptable, too low or too high. ⋯ Pressures generated by anaesthetists inflating tracheal tube cuffs were very variable. Median (IQR [range]) pressures for each pilot balloon ranged from 29 (17-43 [9-56]) cmH(2)O to 74 (49-114 [4-140]) cmH(2)O (p < 0.001). The design of the pilot balloon significantly affects anaesthetists' ability to inflate tracheal tube cuffs to safe pressures.