Anaesthesia
-
Different introducers are available to assist with tracheal intubation. Subtle differences in the design of introducers can have a marked effect on safety and performance. The Difficult Airway Society's Airway Device Evaluation Project Team proposal states that devices should only be purchased for which there is at least a case-control study on patients assessing airway devices. ⋯ Median time to successful placement in the trachea varied from 9 s (Frova) to 22 s (Armstrong) (p = 0.037). We found that the Armstrong and P3 devices were not as acceptable as other introducers and, without significant improvements to their design and characteristics, the use of these devices in studies on patients is questionable. The study protocol is suitable for differentiating between different introducers and could be used as a basis for assessing other types of devices.