[Causes of failure and dangers in the use of motor driven infusion pumps. Accidental closure of the infusion system].
Syringe drivers are used in anaesthetics, intensive care and emergency medicine to deliver small volumes of highly potent drugs with continuous, constant and reproducible flow. For early recognition of interruptions of the drug delivery caused by occlusion of the infusion system, an alarm is triggered as soon as the system pressure exceeds a certain limit. The sensitivity of this alarm depends on the flow rate, type-specific cut-off pressure and the elastic parameters of the infusion system. The sudden release of pressure built up in the system after occlusion occurred can cause delivery of an uncontrolled drug bolus and hence an additional hazard. ⋯ Unidentified occlusions of the system cause grave malfunctioning of syringe drivers. While applying highly potent drugs, the discontinuation of drug delivery with subsequent bolus application can cause vital danger to the patient. As a result of the significant time delays in the pressure alarms, the devices tested do not provide sufficient protection against unrecognized system occlusion. Syringe drivers with adjustable alarm pressure can be set close to the actual infusion pressure. A further important point is that one should aim at a reduction in the elastic properties of the infusion set because of the great impact on alarm delay and bolus volume.