British journal of anaesthesia
-
Comparative Study
Reducing the risk of fatal and disabling hypoglycaemia: a comparison of arterial blood sampling systems.
In 2008, the National Patient Safety Agency (NPSA) published a report after 42 incidents and two deaths where glucose-containing flush solutions were attached to the arterial line. The molar concentration of 5% glucose is 277 mmol litre(-1). Only a tiny amount of sample contamination will lead to an artificially high glucose. As the NPSA sought a solution, a bench model was constructed to compare the performance of three open and three closed arterial line systems in limiting sample contamination. ⋯ Recommended minimal discard volumes are inadequate in the presence of glucose as the flush solution and can lead to high blood glucose readings, inappropriate insulin use, and iatrogenic neuroglycopaenia. Our study demonstrates that the closed-loop arterial sampling system could be the universal solution sought by the NPSA.
-