British journal of anaesthesia
-
Observational Study
Limits of agreement between measures obtained from standard laboratory and the point-of-care device Hemochron Signature Elite(R) during acute haemorrhage.
Rapid diagnosis of coagulopathy in the bleeding patient using point-of-care (POC) devices would be ideal. The Hemochron Signature Elite(®) (HC(®)) is a POC device that determines international normalized ratio (INR) and activated partial thromboplastin time (aPTT). The aim of the study was to evaluate the agreement for INR and aPTT between the HC(®) and standard laboratory values in acute haemorrhage. ⋯ The results showed a lack of agreement between the INR-HC(®) and the aPTT-HC(®) measurements and the standard laboratory in the context of acute haemorrhage. The INR-HC(®) showed moderate performance as a decision-making tool to detect coagulopathy in the context of acute haemorrhage.